飛利浦BV Libra C臂故障信息查詢
error | displayed STUIF texts | LOG file text and remarks (co_error.c): |
019 | X-ray generator not available | Invalid error id or no error message defined !!! |
020 | Warning: system parameters | All DATABASE data has been reset or new DATABASE has |
021 | Warning: system parameters | Configuration data has been reset. The default settings are now |
022 | Warning: system parameters | No BV-scope calibration has been performed, or calibration data |
023 | Warning: system parameters | The APF data has been reset. The default settings are now used. |
024 | Warning: system parameters | The heat integrator data has been reset. This will correct itself |
025 | Warning: system parameters | The current patient dose data has been reset. No action required. |
026 | Warning: system parameters | Annabelle DATABASE data has been reset. The default settings |
027 | Warning: system parameters | The dose output calibration data has been reset. Perform the |
028 | Warning: system parameters | The tank history data has been reset. NOTE: The old tank history |
029 | Warning: system parameters | The energy storage history data has been reset. NOTE: The old |
030 | Warning: system parameters | NOTE: The dose report of the current acquisition patient is |
031 | Warning: system parameters | NOTE: The filament adjustment data has been reset. Perform the |
<none> | Mobile Viewing Station not | <none> |
<none> | Mobile Viewing Station in | <none> |
050 | System initialization problem | A system startup problem occurred. Switch the system off and |
error | displayed STUIF texts | LOG file text and remarks (co_error.c): |
051 | Software version mismatch | The software version installed on the stand is not compatible with |
052 | System not available | The video frequency of the stand is not compatible with the video |
053 | Subsystem error, system still | The DFI is not responding within the allowed time, causing a |
054 | APF parameters not | The APF data is invalid due to an incompatibility between the |
055 | APF parameters not | Unknown IP set. Check the application data for errors. Load DFI |
<none> | Storage speed not accepted Storage speed not selectable | The selected frame speed is not acceptable during X-ray |
056 | Subsystem error, system still | The collimator control board is not responding within the allowed |
057 | Subsystem error, system still | The generator control board is not responding within the allowed |
058 | Subsystem error, system still | The TVI sub-system is not responding within the allowed time, |
059 | Subsystem error, system still | The pulse master control is not responding within the allowed |
060 | HIP time exceeded, low dose | Maximum hip time reached so hip switched off. |
061 | Warning: wrong Viewing | Incorrect DFI connected with the SUCO. Find the correct DFI and |
062 | Image synchronization | No sync detected between XTV and DFI. Check XTV camera and |
<none> | Dose limitation active! only | Dose limitation active, only low dose fluoro. Reduced |
<none> | Dose limitation active! kV | Dose limitation active, kV limited. Reduced performance. |
<none> | Press H/F switch! | <none> |
<none> | Keep H/F switch pressed ... | <none> |
<none> | Adjustment in progress ... | <none> |